Customization: | Available |
---|---|
Auto/Semi-Auto: | Automatic Testing |
Testing Speed: | 6-8PCS/Min |
Suppliers with verified business licenses
Testing Machine | Parameter value |
External Dimension | 1100mm*1240mm*1950mm |
Speed | 6-8pcs/min |
Feeding Mode | Manual |
Input Pressure | ≥0.6MPa |
Input Voltage | AC 220V 50Hz |
Cleanliness | <100,000 class in equipment |
Data Detection Method | Pressure sensor,visual inspection,electronic scales |
Sensor Basic Accuracy | 0.2N |
Vision Basic Accuracy | 0.02mm |
Basic Precision of Electronic Scales | 0.001g |
Software Control Methods | PC |
Software Human Machine Interaction | Industrial controller |
Product Information Recording Methods | Scanner |
Operating Temperature | 20±5°C |
Operating Relative humidity | 30-60 percent non-condensation |
Air Environment | No dust,flammable and explosive,corrosive gases |
Working System | Progression |
Safety Protection | Safety door lock+dual start button+safety light curtain+emergency stop switch |
Test Item | Description | Standard | Accuracy | Reference Standards |
Cap Removal Force | Minimum force required to remove the pen cap | 6-55N | 0.2N | ISO 14971, YBB40012012 |
Activation Force | Minimum force required to press and start the injection process | 2-15N | 0.2N | ISO 11608-5, ISO 14971 |
Trigger Force | Force required to initiate injection | 2-15N | 0.2N | ISO 11608-5, ISO 14971 |
Holding Force | Force to maintain the injection process | 2-15N | 0.2N | ISO 11608-5, ISO 14971 |
Rebound Force | Inner shell rebound force after injection completed | > 0N | 0.2N | ISO 11608-5, ISO 14971 |
Defeat Force | Minimum force required to break the needle cap protection device through vertical pressing | > 48N | 0.2N | ISO 11608-5, ISO 14971, ISO 23908 |
Injection Volume | Weight of collected liquid after injection | > 1 ml | 0.001g | ISO 11608-1, ISO11608-5, YBB40012012, ISO 14971 |
Injection Time | Time interval from plunger start to stop moving | 5-10s | 0.04s | MDER 2015:8 437-484 |
Needle Penetration Depth | Injection depth | 4-8 mm | 0.02 mm | ISO 11608-5, ISO 14971 |
Needle Safety Distance | Needle tip protection distance after injection | ≥ 3.7 mm | 0.02 mm | ISO 11608-5, ISO 23908 |